Metrohm TitrIC 2 User Manual

Installation instructions of titric 2, Application bulletin 286 e, Metrohm

background image

Application Bulletin 286 e

Version 1.0

Installation Instructions of TitrIC 2

861 Dual Channel Advanced Compact IC with 855 Robotic
Titrosampler

VENT

20

mL

800 Dosino

Metrohm

855

VENT

10

mL

800 Dosino

Metrohm

VENT

50

mL

800 Dosino

Metrohm

HNO

3

712

3

x

800

2.861.0010

2.861.0020

The TitrIC 2 System with two 861 Compact ICs and an 855 Robotic Titrosampler

The TitrIC 2 System is used for the fully automatic analysis of water samples using direct measure-
ment, titration and ion chromatography. The following measured values are determined in a very short
time: temperature, conductivity, pH, acid capacity and, in parallel, the concentrations of the individual
anions and cations. Further Metrohm instruments can be incorporated in the existing system at any
time and used to measure additional parameters.

Analytical sequence

1. Conductivity is measured in the sample vessel on the sample changer.
2. First sample transfer to the two 861 Advanced Compact ICs.
3. The IC measurements are started automatically.
4. Second sample transfer to the external cell of the 855 Robotic Titrosampler.
5. Temperature, pH and acid capacity (p and m values) are determined.

The whole procedure is controlled by the TitrIC software: The user enters the sample position and
sample identification, IC system and tiamo method, after which TitrIC transfers all the relevant
information to tiamo and IC Net. When the measurement has been concluded, TitrIC automatically
generates a joint report containing all results. This report can be printed out or produced as an Excel
table.