6. Cleaning

6.3013.XXX and 6.3014.XXX Exchange Units, 8.110.1346

39

6. Cleaning

The Exchange Unit has to be emptied and cleaned only if the

reagent caused any precipitation or if you wish to fill the Ex-

change Unit with a different reagent. Proceed as follows:

1.

Empty the Exchange Unit as much as possible.

2.

Set Dosimat/Titrino in zero position.

3.

Remove the tubing connections of the reagent bottle and of

the buret tip.

4.

Remove light protection (thermostat jacket).

5.

Undo the screw nut of the glass cylinder.

6.

Empty cylinder completely using the 6.2739.010 key and

carefully pull out the piston.

7.

Clean all parts. Take care that no reagent remains in the

threaded holes.

8.

Lightly grease piston at the side with 6.2803.010 grease.

Wipe off leading edge to ensure that the reagent does not

come into contact with the grease.

Do not grease piston for pipetting.

9.

Reassemble Exchange Unit.

Tighten the tubing connection first at the glass cylinder, then

at the stopcock.

7. Accuracy

Accuracy data:

Buret volume V

nominal

mL

Abs. error rel. to

V

nominal

mL

Reproducibility

mL

1

0.003

0.001

5

0.015

0.005

10

0.02

0.01

20

0.03

0.01

50

0.05

0.04

Checking accuracy:

Dosing accuracy - Application Bulletin No. 238

Potentiometric titrations - Application Bulletin No. 252

KF titrations - Application Bulletin No. 255