4. Stopcock

6.3013.XXX and 6.3014.XXX Exchange Units, 8.110.1346

37

4. Stopcock

Two stopcocks are available: The 6.1542.020 PCTFE/PTFE

stopcock and the 6.1542.010 ceramics stopcock.

Use the ceramics stopcock if hard crystals can precipitate from

your solution or if your dosing is continuously running. If the pre-

cipitate is soft and if you use the Exchange Unit at greater time

intervals, we recommend the PCTFE/PTFE stopcock.

PCTFE/PTFE stopcocks are worn out after a certain time. They

have to be replaced more frequently than ceramics stopcocks.

For the following reagents and applications, we recommend:

6.1542.020

PCTFE/PTFE

6.1542.010

ceramics

Acids, aqueous

l

l

Alkali, aqueous

l

Continuous running

l

EDTA

l

HClO

4

in acetic acid

l

l

Iodine, solution

l

Karl Fisher reagent

l

KOH in ethanol

l

Organic solvents

l

Permanganate, KMnO

4

l

Silver nitrate, AgNO

3

l

l

TBAOH

l

Thiosulphate, Na

2

S

2

O

3

l

l

What can you do with a blocked stopcock?

1. Remove stopcock from the handle of the Exchange Unit.

2. Soak stopcock in appropriate solvent. Use

hot water for aqueous solutions;

the corresponding organic solvent for non-aqueous solutions;

methanol, than water, than methanol for KF reagent.

3. If you can turn the stopcock, remount it onto the Exchange

Unit.

You may open the stopcock opening the screw nut. The polished surfaces

of the cock plates must be without damage.

Mount the dried parts and do not touch the polished surface of the cock

plates.