Metrohm Additional eluent for 941 User Manual
Page 17

■■■■■■■■■■■■■■■■■■■■■■
3 Installation
IC equipment: Additional eluent for 941
■■■■■■■■
11
3 Connecting the Dosino to the ion chromatograph
NOTE
The 941 Eluent Production Module must be switched off when
the Dosino is being plugged to the MSB connector.
■
Check whether the 941 Eluent Production Module is switched on.
If so, switch off the instrument.
■
Plug the Dosino cable into one of the 941 Eluent Production
Module's MSB sockets.
Installing the FEP tubings on the Dosino
Accessories
For this step you need the following accessories:
■
800 Dosino (2.800.0010) with 807 Dosing Unit 50 mL (6.3032.250)
attached
■
The three FEP tubings connected to the bottles (6.1805.530),
one additional FEP tubing (6.1805.530)
■
Adapter Dosino Port 4 (6.1808.280)
■
Adapter UNF / M6 (6.2744.080)
1 Connecting the lateral ports
1
Concentrate
2
Port 1
3
Port 3
4
Eluent
■
Screw the concentration bottle's FEP tubing onto Port 1 of the
Dosino.
■
Screw the eluent bottle's FEP tubing onto Port 3 of the Dosino.