Page 14

3.3 Connecting the ultrapure water canister
■■■■■■■■■■■■■■■■■■■■■■
8
■■■■■■■■
IC equipment: Additional eluent for 941
Figure 5
Installing the concentrate bottle cap without level sensor
1 Connecting the FEP tubing
Fasten the tubing connection to the Dosino to the M6 opening.
■
Insert the FEP aspiration tubing into the M6 opening.
■
Screw the FEP tubing into the M6 opening.
2 Sealing the M8 opening
Screw the M8 threaded stopper into the M8 opening.
3 Mounting the adsorber tube
Fill the adsorber tube with adsorber material and cotton (for anion
eluents) or only with cotton (for cation eluents) and insert it into the
large opening. Secure in place with an SGJ clip.
3.3
Connecting the ultrapure water canister
Assembling the bottle cap without level sensor
The canister with the ultrapure water can be equipped with a bottle cap
without level sensor, e.g. with the eluent bottle cap (6.1602.160).
Accessories
For this step you need the following accessories:
■
Canister, 10 L (6.1621.000)
■
Eluent bottle cap (6.1602.160)
■
FEP aspiration tubing (6.1819.110)
■
FEP tubing (6.1805.530)
■
Adsorber tube (6.1609.000), included in the accessories for the dosing
unit (6.3032.250)