2 protecting the separation column, 3 storing the separation column, 4 regenerating the separation column – Metrohm 930 Compact IC Flex SeS User Manual
Page 93: Protecting the separation column, Storing the separation column, Regenerating the separation column
■■■■■■■■■■■■■■■■■■■■■■
5 Operation and maintenance
930 Compact IC Flex SeS (2.930.1400)
■■■■■■■■
85
in the Ion
Chromatography product area. You can find information on special IC
applications in the corresponding "Application Bulletins" or "Applica-
tion Notes", which are available on the Internet at
request from your responsible Metrohm representative.
5.14.2
Protecting the separation column
We recommend using the following protective measures so that the sepa-
ration column retains its separating efficiency for as long as possible:
■
Microfilter both the sample and the eluent (0.45 µm filter) and aspirate
the eluent via the aspiration filter (6.2821.090) as well.
■
Always use a guard column. Information regarding which guard col-
umn is suitable for your separation column can be found in the
Metrohm IC Column Program (which is available from your
Metrohm representative), the leaflet provided along with your separa-
tion column, the product information about the separation column at
(Ion Chromatography product area), or
obtained directly from your representative.
■
Always use the pulsation absorber.
5.14.3
Storing the separation column
Always store separation columns you do not need in a sealed and filled
state according to the column manufacturer's specifications.
5.14.4
Regenerating the separation column
The separation column can be regenerated according to the column man-
ufacturer's specifications if the separation characteristics of the column
have deteriorated. You can find information on regenerating separation
columns available from Metrohm on the leaflet provided with every col-
umn.
NOTE
Regeneration is intended as a last resort. It should not be carried out
regularly.