Figure 7, Installing the eluent bottle cap – Metrohm 930 Compact IC Flex SeS User Manual
Page 28
3.5 Connecting the eluent bottle
■■■■■■■■■■■■■■■■■■■■■■
20
■■■■■■■■
930 Compact IC Flex SeS (2.930.1400)
■
The eluent bottle cap GL 45 accessory set (6.1602.160)
This accessory set contains the bottle cap, an M6 tubing nipple, an M8
tubing nipple, two O-rings and an M6 and M8 threaded stopper.
■
The tubing adapter for aspiration filter accessory set (6.2744.210)
This accessory set contains a filter holder, a clamping screw and tubing
weighting.
■
An aspiration filter (6.2821.090)
Connecting the eluent aspiration tubing
1 Installing the eluent bottle cap (6.1602.160)
■
Start by pushing the M8 tubing nipple onto the loose end of the
eluent aspiration tubing, followed by the O-ring.
■
Push the loose end of the eluent aspiration tubing through the
M8 opening of the bottle cap and screw it on for the time being.
Figure 7
Installing the eluent bottle cap
2 Mounting the tubing adapter and the aspiration filter
Install the parts of the tubing adapter for aspiration filter
(6.2744.210) accessory set:
■
Start by pushing the tubing weighting onto the loose end of the
eluent aspiration tubing.
■
Then push the clamping screw onto the loose end of the eluent
aspiration tubing.
■
Lastly, push the filter holder onto the loose end of the eluent aspi-
ration tubing and screw it onto the tubing nipple.
The end of the tubing should extend approximately 1 cm.
Installing the aspiration filter:
NOTE
Only handle the aspiration filter while wearing gloves.