3.10 Suppressor
■■■■■■■■■■■■■■■■■■■■■■
34
■■■■■■■■
930 Compact IC Flex SeS (2.930.1400)
1 Fasten the capillary labeled regenerant to the outlet of the inline filter
(6.2821.120) using a pressure screw (6.2744.070).
2 Use a pressure screw (6.2744.070) to fasten one end of the PEEK
capillary (6.1803.030) to the inlet of the inline filter.
3 Use a pressure screw (6.2744.070) and an adapter (6.2744.080) to
fasten the other end of the PEEK capillary to Port 1 of the Dosino.
4 Connect the FEP tubing (6.1805.120) from the bottle with the regen-
eration solution to Port 2 of the Dosino.
3.10.2.3
Connecting the rinsing solution
There are 2 options for rinsing the suppressor:
■
Rinsing solution via STREAM (recommended)
Use the eluent from the conductivity detector as a rinsing solution.
■
Rinsing solution via Dosino
Prepare the rinsing solution in its own bottle and transport with a Dos-
ino.
The rinsing solution is connected to the rinsing solution capillary.
- 915 KF Ti-Touch (382 pages)
- 800 Dosino (53 pages)
- 767 Calibrated Reference (23 pages)
- 940 Professional IC Vario ONE/SeS/Prep 2 (54 pages)
- 754 Dialysis Unit (49 pages)
- 815 Robotic Soliprep for LC (76 pages)
- Vision Manual (207 pages)
- tiamo 2.1 Manual (1532 pages)
- 825 Lab Link (37 pages)
- 808 Titrando (70 pages)
- 902 Titrando (52 pages)
- 756 KF Coulometer (163 pages)
- 756 KF Coulometer (162 pages)
- 940 Professional IC Vario ONE/LPG (98 pages)
- 850 Professional IC Anion MCS Prep 3 (154 pages)
- 850 Professional IC Anion MCS Prep 3 (152 pages)
- 904 Titrando (58 pages)
- 850 Professional IC Anion MSM-HC MCS Prep 2 (150 pages)
- 930 Compact IC Flex Oven/ChS/Deg (47 pages)
- 872 Extension Module Liquid handling (64 pages)
- 814 USB Sample Processor (90 pages)
- 814 USB Sample Processor (91 pages)
- 940 Professional IC Vario (43 pages)
- Vision – Tutorial (40 pages)
- 799 GPT Titrino (242 pages)
- 889 IC Sample Center (68 pages)
- 761 Compact IC (228 pages)
- 851 Titrando (100 pages)
- 748 DH Sample Changer (32 pages)
- 940 Professional IC Vario ONE/SeS/HPG (51 pages)
- 896 Professional Detector – Amperometry (62 pages)
- 877 Titrino plus (139 pages)
- 881 Compact IC pro – Anion (129 pages)
- 940 Professional IC Vario ONE/ChS/HPG (112 pages)
- 930 Compact IC Flex Deg (41 pages)
- 840 PC Control 5.0 / Touch Control (351 pages)
- 940 Professional IC Vario ONE/Prep 1 (45 pages)
- 776 Dosimat (42 pages)
- 717 Sample Changer (36 pages)
- 815 Robotic USB Sample Processor XL (113 pages)
- 815 Robotic USB Sample Processor XL (114 pages)
- 940 Professional IC Vario ONE/SeS/PP (126 pages)
- 838 Advanced Sample Processor Installation Instructions (109 pages)
- 700 Dosino (55 pages)
- 719 S Titrino (152 pages)