7.1 Reference conditions
■■■■■■■■■■■■■■■■■■■■■■
44
■■■■■■■■
2.896.0030 896 Professional Detector – Conductivity & Amperometry
7 Technical specifications
7.1
Reference conditions
The technical data listed in this Chapter refers to the following reference
conditions:
Ambient tempera-
ture
+25 °C (± 3 °C)
Instrument status
> 40 minutes in operation (equilibrated)
7.2
Conductivity detector
Type
■
Microprocessor-controlled Digital Signal Processing (DSP technol-
ogy)
■
Intelligent detector with 6 reference chromatograms
Measuring range
0…15000 µS/cm without range switching
Noise
< 0.1 nS at 1 µS/cm
Deviations from
the linearity
■
< 0.1 % for conductivity values higher than 16 µS/cm
■
< 1 % for conductivity values lower than 16 µS/cm
Drift
< 0.2 nS/cm per hour
Measuring rate
10 measurements per second for optimum results without filtering
Resolution
0.0047 nS/cm
Baseline
Noise < 0.2 nS/cm typical for sequential suppression
Conductivity
detector
Cell volume
0.8 µL
Cell constant
■
Individual calibration data saved in the detector
■
adjustable in the range: 13.0…21.0 /cm
Electrodes
Ring-shaped electrodes made from stainless steel
Materials in
contact with
eluent
Chemically inert PCTFE
Maximum oper-
ating pressure
5.0 MPa (50 bar)
Cell tempera-
ture
20…50 °C in steps of 5 °C