4.7 Connecting the electrode connection cables
■■■■■■■■■■■■■■■■■■■■■■
28
■■■■■■■■
2.896.0030 896 Professional Detector – Conductivity & Amperometry
2 Remove the purge stopper
Remove the stopper from the Purge connector.
3 Deaerate the measuring cell
Observe the eluent that emerges through the deaeration opening.
Wipe up liquid with a cloth.
Once no more air bubbles are visible, screw the stopper back on the
Purge connector and tighten it by hand.
4 Switch off the high pressure pump in MagIC Net™.
4.7
Connecting the electrode connection cables
Connecting the electrode connection cables to the detector
Precondition: The electrode connection cables are connected to the meas-
uring cell (see chapter installation in the manual of the measuring cell).
1 Plug in the straight plug of the working electrode connection cable
(red sleeve) into the WE socket of the detector.
2 Plug in the straight plug of the reference electrode connection cable
(black sleeve) into the RE socket of the detector.
3 Plug in the straight plug of the auxiliary electrode connection cable
(blue sleeve) into the AE socket of the detector.