4 safety notes, 1 electrical safety, 2 general precautionary rules – Metrohm 752 Pump Unit User Manual
Page 9

1 Introduction
752 Pump Unit
6
1.4
Safety notes
1.4.1
Electrical safety
While electrical safety in the handling of the 752 Pump Unit is assured
in the context of the specifications IEC 1010-1 (protection class 1, de-
gree of protection IP40), the following points should be noted:
•
Mains connection
Setting the mains voltage, checking the mains fuse and the mains
connection must be effected in accordance with the instructions in
section 2.2.
•
Opening the 752 Pump Unit
When the 752 Pump Unit is connected to the power supply the in-
strument must neither be opened nor should parts be removed from
it, otherwise there is a danger of coming into contact with compo-
nents which are live. Always disconnect the instrument from all voltage
sources before you open it and ensure that the mains cable is
disconnected from mains connection 15
15 !
•
Protection against static charges
Electronic components are sensitive to static charging and can be
destroyed by discharges. Before you touch any of the components
inside the 752 Pump Unit, you should earth yourself and any tools you
are using by touching an earthed object (e.g. housing of the instru-
ment or a radiator) to eliminate any static charges which exist.
1.4.2
General precautionary rules
•
Handling of solvents
Check all pump tubings and all input and output lines periodically for
possible leaks. Follow the relevant instructions regarding the handling
of flammable and/or toxic solvents and their disposal.
•
Regular exchange of pump tubings
Pump tubings are consumable material and must be replaced from
time to time (see section 4.4). Suitable measures must be taken so
that any leak which might occur in the pump tubing or connections
during unattended operation will cause no damage (placing the
instrument at the bottom, collection device for any liquid which may
leak out).