16 switch on remote control, Switch pump on/off in a program, Switch pump on/off at an event – Metrohm 752 Pump Unit User Manual
Page 19

3 Operation
752 Pump Unit
16
Switch on remote control
ON
OFF
REMOTE
In order to allow remote operation via 732 IC Detector the remote
control must be switched on.
ON
Remote control switched on
OFF
Remote control switched off
Automatic pump switch-on when the 732 IC
Detector is switched on
In order to start up the drive of the 752 Pump Unit automatically
when the 732 IC Detector is switched on alterations must be
made to the basic settings (Setup, see section 4.4.1 of 732/733
Manual). The remote output lead 1 must be set to 1 (on, active,
0 V):
>SETUP/output
remote 10000000
When the 732 IC Detector is switched off the pump is also
switched off.
Switch pump on/off in a program
In the 732 IC Detector the
create time programs with a maximum of 20 program steps (see
section 4.7.1 of 732/733 Manual). The pump can be switched on
or off at any program step by setting the remote output lead 1 to
1 (on, active, 0 V) or 0 (off, inactive, open) respectively:
Switch on:
>PROGRAM/edit XXX.X min
remote
1
TTTTTTT
Switch off:
>PROGRAM/edit XXX.X min
remote
0
TTTTTTT
Switch pump on/off at an event
In the 732 IC Detector the
program a maximum of 4 different events (see section 4.7.3 of
732/733 Manual). The pump can be switched on or off at any
event by setting the remote output lead 1 to 1 (on, active, 0 V) or
0 (off, inactive, open) respectively:
Switch on:
EVENT YY-MM-DD HH:MM:SS
remote
1
TTTTTTT
Switch off:
EVENT YY-MM-DD HH:MM:SS
remote
0
TTTTTTT