2.3 Connection to the IC System

752 Pump Unit

13

Attach inlet capillary 90

90 using compression fitting 20

20

(6.2744.014) to the connector 23

23 (see Fig. 6) of the filter unit

PEEK (see section 2.6.3 of 732/733 Manual).

Attach a piece of PTFE tubing 16

16 cut to the required length

using a compression fitting 20

20 (6.2744.014) to the connector

21

21 with filter at the other end of the filter unit PEEK.

Attach the other end of the PTFE tubing 16

16 using a compres-

sion fitting 20

20 (6.2744.014) to the coupling 17

17 (6.2744.030)

and mount it on the outlet end of the first length of pump

tubing 18

18.

Mount a coupling 17

17 (6.2744.030) to the inlet end of the first

length of pump tubing 18

18. Attach a piece of the PTFE tubing

16

16 (6.1803.020) cut to the required length using a compres-

sion fitting 20

20 (6.2744.014) to the other end of this coupling.

Immerse the other end of the aspirating tubing in a vessel

containing regeneration solution (normally 20 mmol/L H

2

SO

4

)

and fix in place.

Pull outlet capillary 91

91 marked with "Waste" of the suppressor

module through the rear panel opening 43

43, lead it to a suffi-

ciently large waste container and fix it in place.

9 Suppressor connection 3: H

2

O

Loosen rotary nipple screwed onto the interior side of con-

nection 28

28. Pull inlet capillary 93

93 marked with "H

2

O" (see

Fig. 17 of 732/733 Manual) by hand out of the opening of

connection 28

28 as far as required. Retighten nipple on the in-

terior side of connection 28

28 to fix inlet capillary 93

93.

Attach inlet capillary 93

93 using compression fitting 20

20

(6.2744.014) to the connector 23

23 (see Fig. 6) of the second

filter unit PEEK (see section 2.6.3 of 732/733 Manual).

Attach a piece of PTFE tubing 16

16 cut to the required length

using a compression fitting 20

20 (6.2744.014) to the connector

21

21 with filter at the other end of the filter unit PEEK.

Attach the other end of the PTFE tubing 16

16 using a compres-

sion fitting 20

20 (6.2744.014) to the coupling 17

17 (6.2744.030)

and mount it on the outlet end of the second length of pump

tubing 18

18.

Mount a coupling 17

17 (6.2744.030) to the inlet end of the

second length of pump tubing 18

18. Attach a piece of the PTFE

tubing 16

16 (6.1803.020) cut to the required length using a

compression fitting 20

20 (6.2744.014) to the other end of this

coupling.

Immerse the other end of the aspirating tubing in a vessel

containing rinsing solution (normally dist. H

2

O) and fix in

place.

Pull outlet capillary 92

92 marked with "Waste" of the suppressor

module through the rear panel opening 43

43, lead it to a suffi-

ciently large waste container and fix it in place.