2 Installation

752 Pump Unit

14

10 Startup of 752 Pump Unit

Set switch 22 on 752 Pump Unit to "OFF".

Switch on 752 Pump Unit with mains switch 12

12 and set switch

33 to "ON".

Adjust the contact pressure for both tubing cartridges: press

contact pressure lever 55 upwards until the solutions just start

to be drawn in. Then press the contact pressure lever up-

wards until it clicks once more to obtain optimal contact pres-

sure.

Check all capillaries between the 752 Pump Unit, the sup-

pressor module and the waste container for escaping liquid. If

liquid escapes anywhere the corresponding compression fit-

ting must be tightened further or changed.

Pump tubings are consumable material with a lifetime which depends

on the contact pressure. This is why the tubing cartridges should be

raised completely by loosening snap-action lever 77 on the right-hand

side if the pump is to remain switched off for a considerable length of

time (the set contact pressure remains unchanged).