beautypg.com

Metrohm 752 Pump Unit User Manual

Page 14

background image

2.3 Connection to the IC System

752 Pump Unit

11

1

3

2

89

89

92

92

93

93

94

94

90

90

91

91

Fig. 5:

Connections at suppressor module

3 Connect column to suppressor module

Cut inlet capillary 89

89 marked with "Eluent" of suppressor

module 95

95 (see Fig. 17 in 732/733 Manual) to the desired

length using a sharp cutting tool (e.g. razor blade).

Screw inlet capillary 89

89 to outlet end of separating column 76

76

using a 6.2744.014 Compression fitting.

4 Fix column

Insert one or two column holders 84

84 (6.2027.030, 6.2027.040

or 6.2027.050) in mounting rails 83

83 and fasten separating

column in the column holder.

5 Connect suppressor module to detector block

Cut outlet capillary 94

94 marked with "Detector" of suppressor

module 95

95 to the desired length using a sharp cutting tool

(e.g. razor blade).

Screw outlet capillary 94

94 marked with "Detector" to coupling

96

96 using a 6.2744.014 Compression Fitting.

Screw inlet capillary 82

82 of detector block 81

81 to other end of

coupling 96

96.

6 Fix connection suppressor – detector block

Insert one of the column holders 84

84 (6.2027.030, 6.2027.040

or 6.2027.050) in mounting rail 83

83 and fasten coupling 96

96 in

the column holder.

7 Attach pump tubings to 752 Pump Unit

Loosen both tubing cartridges 44 from the holding clamp 66 by

pressing down snap-action lever 77 and remove from mount-

ing pin 99 on Pump Unit 752 (see Fig. 1).

89

89

Suppressor inlet

capillary for eluent

90

90

Suppressor inlet

capillary for H

2

SO

4

91

91

Suppressor outlet

capillary for H

2

SO

4

92

92

Suppressor outlet

capillary for H

2

O

93

93

Suppressor inlet

capillary for H

2

O

94

94

Suppressor outlet

capillary for eluent

Eluent

H

2

O

Detector

Waste

Waste

H

2

SO

4