Figure 4, Disassembling the measuring cell – Metrohm Wall-Jet-Cell User Manual
Page 22
![background image](/manuals/586500/22/background.png)
5.2 Maintenance
■■■■■■■■■■■■■■■■■■■■■■
16
■■■■■■■■
IC equipment Wall-Jet cell
■
The measuring cell has been removed from the detector.
■
The electrode cables have been removed.
To disassemble the measuring cell, you need a 2.5 mm hex key.
1
2
3
4
Figure 4
Disassembling the measuring cell
1
Fastening screw (4x)
2
Base part
3
Spacer
6.1257.810 or 6.1257.830.
4
Measuring part
With auxiliary electrode.
1 Remove the reference electrode.
2 Remove the working electrode.
3 Unscrew the four screws on the base part of the measuring cell with
the hex key and remove them.
4 Remove the base part.
5 Remove the spacer.
Replace the spacer as follows: