3 Installation

IC equipment Wall-Jet cell

■■■■■■■■

9

2 Connecting the cell outlet

Use a pressure screw (6.2744.014) to fasten a piece of PEEK capillary
(6.1831.010) with a length of 1 to 1.5 m to the Out connector of
the measuring cell.

NOTE

This capillary must be tested with the IC instrument the first time
the detector is put into operation, see the chapter "Start-up" in
the manual for the detector.

If the preheating capillary is used, proceed as follows:

Connecting capillaries to the measuring cell

1 Connecting the preheating capillary

NOTE

The preheating capillary must be tested with the IC instrument the
first time the detector is put into operation, see the chapter "Start-
up"
in the manual for the detector.

Use a pressure screw (6.2744.014) to connect the detector input
capillary to the Eluent in connector of the detector.

Use a pressure screw (6.2744.014) to fasten a piece of PEEK capil-
lary (6.1831.010) to the Eluent to cell connector of the detec-
tor.

2 Connecting the cell inlet

Use a pressure screw (6.2744.014) to fasten the other end of the
PEEK capillary (6.1831.010) to the In connector of the measuring
cell.

3 Connecting the cell outlet

Use a pressure screw (6.2744.014) to fasten a piece of PEEK capillary
(6.1831.010) with a length of 1 to 1.5 m to the Out connector of
the measuring cell.