3 Installation
IC equipment Wall-Jet cell
■■■■■■■■
9
2 Connecting the cell outlet
Use a pressure screw (6.2744.014) to fasten a piece of PEEK capillary
(6.1831.010) with a length of 1 to 1.5 m to the Out connector of
the measuring cell.
NOTE
This capillary must be tested with the IC instrument the first time
the detector is put into operation, see the chapter "Start-up" in
the manual for the detector.
If the preheating capillary is used, proceed as follows:
Connecting capillaries to the measuring cell
1 Connecting the preheating capillary
NOTE
The preheating capillary must be tested with the IC instrument the
first time the detector is put into operation, see the chapter "Start-
up" in the manual for the detector.
■
Use a pressure screw (6.2744.014) to connect the detector input
capillary to the Eluent in connector of the detector.
■
Use a pressure screw (6.2744.014) to fasten a piece of PEEK capil-
lary (6.1831.010) to the Eluent to cell connector of the detec-
tor.
2 Connecting the cell inlet
Use a pressure screw (6.2744.014) to fasten the other end of the
PEEK capillary (6.1831.010) to the In connector of the measuring
cell.
3 Connecting the cell outlet
Use a pressure screw (6.2744.014) to fasten a piece of PEEK capillary
(6.1831.010) with a length of 1 to 1.5 m to the Out connector of
the measuring cell.