3 Installation
IC equipment Wall-Jet cell
■■■■■■■■
11
Connecting the electrode cables to the measuring cell
Prerequisites:
■
The working electrode and the reference electrode are inserted into
the measuring cell.
1 Plug the angled plug of the working electrode cable (labeled WE)
into the working electrode socket.
2 Plug the angled plug of the reference electrode cable (labeled RE)
into the reference electrode socket.
3 Plug the angled plug of the auxiliary electrode cable (labeled AE) into
the socket (labeled AE).