Application Bulletin 319 e
Installation Instructions of TitrIC 5
Page 12/16

ƒ At the left click on

Configuration

, then at the top right under

Titrants/Solutions

double-click on the

appropriate (empty) entries and enter the following under

Solution name

and

Concentration

:

Configuration table for titrants/solutions

Now the tubing lengths and ports are defined for each Dosing unit under

Edit/Properties…/ Dosing unit

as

follows:

Anion Cation HCl

Port Length Port Length Port Length

Dosing Port
Prep/Empty

Special Port

-

Dosing Port 1

-

Dosing Port 1

-

Dosing Port 1

Port 4

60 cm

Port 1

60 cm

Port 1

40 cm

Dosing Port 2

Port 3

0 cm

Port 3

0 cm

Port 3

0 cm

Fill Port

Port 2

80 cm

Port 2

80 cm

Port 2

25 cm

Special Port

Port 1

0 cm

Port 4

40 cm

Port 4

0 cm

Configuration table of the three dosing units

6.2.6 Electrodes

ƒ pH electrode: Under

Configuration/Sensors

click on

Edit/New/pH electrode

and enter under

Sensor

name Aquatrode Plus

, type in the desired sensor information (e.g. serial number) and click on

.

ƒ Conductivity sensor: Under

Configuration/Sensors

click on

Edit/New/other sensor

and enter under

Sensor name Conductivity sensor

type in the desired sensor information and click on

.

6.2.7 Methods

The following six methods are provided for tiamo

TM

:

ƒ

TitrIC 5 - preparation of titration equipment:

This method is used for preparing all titration instruments,

i.e., rinsing the tubing connections of the Dosing Units with reagents or deionized water and
rinsing the external cell. Start this method only from tiamo

TM

– not from TitrIC!

ƒ

TitrIC 5 - complete run:

With this method all available parameters are measured and the IC

determination is initiated.

ƒ

TitrIC 5 - IC calibration - anion & cation:

This method is used for calibrating the IC measurements

with standard solutions without the titration part of a typical TitrIC measurement; this means that
a lot of standard solution can be saved.

ƒ

TitrIC 5 – manual pH calibration:

Here the Aquatrode is taken out of the titration vessel and placed

into the buffer solutions. Start this method only from tiamo

TM

!

ƒ

TitrIC 5 – semi automated pH calibration:

With this method the pH electrode is calibrated

automatically; the buffer solutions are flushed into the titration vessel. Start this method only
from tiamo

TM

– not from TitrIC!

ƒ

TitrIC 5 - titer determination:

This method is used for the titer determination. Start this method only

from tiamo

TM

!

ƒ

TitrIC Calibrate 856 conductivity module

This method is used for calibration of the conductivity

module. Start this method only from tiamo

TM

!