Application Bulletin 319 e
Installation Instructions of TitrIC 5
Page 12/16
At the left click on
Configuration
, then at the top right under
Titrants/Solutions
double-click on the
appropriate (empty) entries and enter the following under
Solution name
and
Concentration
:
Configuration table for titrants/solutions
Now the tubing lengths and ports are defined for each Dosing unit under
Edit/Properties…/ Dosing unit
as
follows:
Anion Cation HCl
Port Length Port Length Port Length
Dosing Port
Prep/Empty
Special Port
-
Dosing Port 1
-
Dosing Port 1
-
Dosing Port 1
Port 4
60 cm
Port 1
60 cm
Port 1
40 cm
Dosing Port 2
Port 3
0 cm
Port 3
0 cm
Port 3
0 cm
Fill Port
Port 2
80 cm
Port 2
80 cm
Port 2
25 cm
Special Port
Port 1
0 cm
Port 4
40 cm
Port 4
0 cm
Configuration table of the three dosing units
6.2.6 Electrodes
pH electrode: Under
Configuration/Sensors
click on
Edit/New/pH electrode
and enter under
Sensor
name Aquatrode Plus
, type in the desired sensor information (e.g. serial number) and click on
.
Conductivity sensor: Under
Configuration/Sensors
click on
Edit/New/other sensor
and enter under
Sensor name Conductivity sensor
type in the desired sensor information and click on
.
6.2.7 Methods
The following six methods are provided for tiamo
TM
:
TitrIC 5 - preparation of titration equipment:
This method is used for preparing all titration instruments,
i.e., rinsing the tubing connections of the Dosing Units with reagents or deionized water and
rinsing the external cell. Start this method only from tiamo
TM
– not from TitrIC!
TitrIC 5 - complete run:
With this method all available parameters are measured and the IC
determination is initiated.
TitrIC 5 - IC calibration - anion & cation:
This method is used for calibrating the IC measurements
with standard solutions without the titration part of a typical TitrIC measurement; this means that
a lot of standard solution can be saved.
TitrIC 5 – manual pH calibration:
Here the Aquatrode is taken out of the titration vessel and placed
into the buffer solutions. Start this method only from tiamo
TM
!
TitrIC 5 – semi automated pH calibration:
With this method the pH electrode is calibrated
automatically; the buffer solutions are flushed into the titration vessel. Start this method only
from tiamo
TM
– not from TitrIC!
TitrIC 5 - titer determination:
This method is used for the titer determination. Start this method only
from tiamo
TM
!
TitrIC Calibrate 856 conductivity module
This method is used for calibration of the conductivity
module. Start this method only from tiamo
TM
!