background image

Application Bulletin 319 e
Installation Instructions of TitrIC 5
Page 10/16

5 Miscellaneous

5.1 Stirrer

The 802 Rod Stirrer is inserted through opening 8 of the external cell and connected to the tower of the
855. The propeller (6.1909.020) is pushed on from below.

5.2 Length of the aspiration tips in the external cell

ƒ Clamp the titration vessel to the titration vessel lid.

ƒ The M8 aspiration tip in opening 1 must reach right to the base of the titration vessel as this is

the only way to ensure that the external cell is cleaned properly.

ƒ The buret tips in openings 3 and 4 should reach the middle of the titration vessel.

5.3 Electrodes

ƒ The pH electrode (Aquatrode Plus with Pt 1000; 6.0257.000) is inserted into opening 9. The

cable is connected to the 855 Robotic Titrosampler at the rear top right. The plug of the gray
cable is connected to the opening marked

Ind.

The red plugs with the adapters 6.2103.140 and

6.2103.130 are inserted in the openings marked

Temp.

If the electrode is not used for some time then it should be returned to its holder and kept in the
storage solution (6.2323.000).

ƒ Conductivity cell (6.0915.100): The measuring cell cable is led through the guide chain of the

855 and the plug is inserted into the opening marked

Cond. Cell

of the 856 Conductometer. Later

(as soon as the titration head has been mounted) the measuring cell with the SGJ sleeve
(6.1236.050) will be inserted in the SGJ 14 opening of the titration head.

ƒ

6 Software

6.1 Installation of the software

ƒ tiamo

TM

2.1 – installation (tiamo

TM

CD): All the standard directories proposed by the program

should be accepted.

ƒ MagIC Net 2.0 – installation (MagIC Net CD): All the standard directories proposed by the

program should be accepted. Restart windows.

ƒ TitrIC 2.0 – installation (TitrIC CD): TitrIC 2.0 must only be installed after tiamo

TM

2.1 and

MagIC Net 2.0 have already been installed. After installation the standard TitrIC directory can
be found under

C:\Program files\ Metrohm\TitrIC

, the TitrIC.exe file under

C:\Program

files\Metrohm\tiamo\bin

. The TitrIC database, reports are found on

C:\Documents and Settings\All

Users\ApplicationData\TitrIC.

Æ All programs should be closed!

6.2 Configuration

of

tiamo

TM

Æ The controller cable (6.2151.000), which is already connected to the 855, is now connected to an
USB interface on the PC. Wait a few seconds. The driver installation for the 855 follows. It’s not
necessary to start windows update, click on

automatic installation of the software

.

Æ The controller cable (6.2151.000), which is already connected to the 856 conductivity module, is now
connected to an USB interface on the PC

Æ tiamo

TM

2.1 is started