Installation Instructions of TitrIC 5
Page 10/16
5 Miscellaneous
5.1 Stirrer
The 802 Rod Stirrer is inserted through opening 8 of the external cell and connected to the tower of the
855. The propeller (6.1909.020) is pushed on from below.
5.2 Length of the aspiration tips in the external cell
Clamp the titration vessel to the titration vessel lid.
The M8 aspiration tip in opening 1 must reach right to the base of the titration vessel as this is
the only way to ensure that the external cell is cleaned properly.
The buret tips in openings 3 and 4 should reach the middle of the titration vessel.
5.3 Electrodes
The pH electrode (Aquatrode Plus with Pt 1000; 6.0257.000) is inserted into opening 9. The
cable is connected to the 855 Robotic Titrosampler at the rear top right. The plug of the gray
cable is connected to the opening marked
Ind.
The red plugs with the adapters 6.2103.140 and
6.2103.130 are inserted in the openings marked
Temp.
If the electrode is not used for some time then it should be returned to its holder and kept in the
storage solution (6.2323.000).
Conductivity cell (6.0915.100): The measuring cell cable is led through the guide chain of the
855 and the plug is inserted into the opening marked
Cond. Cell
of the 856 Conductometer. Later
(as soon as the titration head has been mounted) the measuring cell with the SGJ sleeve
(6.1236.050) will be inserted in the SGJ 14 opening of the titration head.
6 Software
6.1 Installation of the software
tiamo
TM
2.1 – installation (tiamo
TM
CD): All the standard directories proposed by the program
should be accepted.
MagIC Net 2.0 – installation (MagIC Net CD): All the standard directories proposed by the
program should be accepted. Restart windows.
TitrIC 2.0 – installation (TitrIC CD): TitrIC 2.0 must only be installed after tiamo
TM
2.1 and
MagIC Net 2.0 have already been installed. After installation the standard TitrIC directory can
be found under
C:\Program files\ Metrohm\TitrIC
, the TitrIC.exe file under
C:\Program
files\Metrohm\tiamo\bin
. The TitrIC database, reports are found on
C:\Documents and Settings\All
Users\ApplicationData\TitrIC.
Æ All programs should be closed!
6.2 Configuration
of
tiamo
TM
Æ The controller cable (6.2151.000), which is already connected to the 855, is now connected to an
USB interface on the PC. Wait a few seconds. The driver installation for the 855 follows. It’s not
necessary to start windows update, click on
automatic installation of the software
.
Æ The controller cable (6.2151.000), which is already connected to the 856 conductivity module, is now
connected to an USB interface on the PC
Æ tiamo
TM
2.1 is started