Removing the filter

Figure 23

Inline filter – removing the filter

1

Filter housing
Inline filter housing. Part of the accessories
(6.2821.120).

2

Filter screw
Screw for the inline filter. Part of the acces-
sories (6.2821.120).

3

Filter (6.2821.130)
Packaging contains 10 pieces.

1 Shutting off the flow

Switch off the high-pressure pump in the software.

2 Removing the inline filter

Unscrew the both pressure screws from the inline filter.

3 Unscrewing the filter screw

Use two adjustable wrenches (6.2621.000) to loosen the filter screw
(23-2) from the filter housing (23-1) and unscrew it by hand.

4 Removing the filter