Figure 8, Installing tubing weighting and aspiration filter – Metrohm 940 Professional IC Vario ONE/ChS/HPG User Manual
Page 31
■■■■■■■■■■■■■■■■■■■■■■
3 Installation
940 Professional IC Vario ONE/ChS/HPG (2.940.1240)
■■■■■■■■
23
2 Mounting the tubing adapter and the aspiration filter
Install the parts of the tubing adapter for aspiration filter
(6.2744.210) accessory set:
■
Start by pushing the tubing weighting onto the loose end of the
eluent aspiration tubing.
■
Then push the clamping screw onto the loose end of the eluent
aspiration tubing.
■
Lastly, push the filter holder onto the loose end of the eluent aspi-
ration tubing and screw it onto the tubing nipple.
The end of the tubing should extend approximately 1 cm.
Installing the aspiration filter:
NOTE
Only handle the aspiration filter while wearing gloves.
■
Place the loose end of the eluent aspiration tubing into the aspira-
tion filter.
The end of the tubing should reach approximately to the center of
the aspiration filter.
■
Tighten the aspiration filter to the filter holder.
Figure 8
Installing tubing weighting and aspiration filter
3 Installing the eluent bottle cap on the eluent bottle
■
Insert the eluent aspiration tubing into the eluent bottle
(6.1608.070).
■
Tighten the bottle cap on the eluent bottle.
■
Adjust the length of the eluent aspiration tubing so that the aspi-
ration filter is at the bottom of the eluent bottle. Then fasten it in
place using the M8 tubing nipple.
■
Seal the M6 opening on the bottle cap with the M6 threaded
stopper from the accessory set.