3.4 Removing transport locking screws

■■■■■■■■■■■■■■■■■■■■■■

18

■■■■■■■■

940 Professional IC Vario ONE/ChS/HPG (2.940.1240)

3.4

Removing transport locking screws

To avoid damage to the drives for the high-pressure pump and the vac-
uum pump during transport, the pumps are secured with transport lock-
ing screws. These are located at the rear of the instrument and labeled
with Transport security screws.

Remove these transport locking screws before the initial start-up.

Accessories

For this step you need:

Hex key 4 mm (6.2621.030)

Remove the transport locking screws

Figure 6

Removing the transport locking screws

1

Transport locking screws
For the vacuum pump.

2

Transport locking screws
For the high-pressure pump.

3

Transport locking screws
For an additional high-pressure pump in the
bottom drawer.

1 Remove all of the transport locking screws with the hex key.