beautypg.com

Metrohm 850 Professional IC Anion MSM-HC MCS LP-Gradient User Manual

Page 24

background image

3.2 Initial installation

■■■■■■■■■■■■■■■■■■■■■■

14

■■■■■■■■

850 Professional IC–Anion – MSM-HC – MCS – LP Gradient

2 Installations on the rear of the instrument

Remove handle and rollers (see Chapter 3.7.1, page 22).

Place the detector in the instrument and connect it (see Chapter
3.7.2, page 25)
.

Remove transport locking screws (see "Removing transport lock-
ing screws", page 27)
.

Connect the leak sensor (see "Connecting the leak sensor", page
27)
.

Connect drainage tubings (see "Installing drainage tubing", page
30)
.

3 Installing the eluent path

Equip the eluent aspiration tubing and connect with the eluent
bottle (see Chapter 3.9.1, page 33).

Instead of the column, connect the 6.2744.040 coupling with a
6.2744.010 pressure screw at the end of the pre-installed column
inlet capillary.

Use a 6.2744.014 pressure screw to connect the capillary of the
MSM-HC labeled with Eluent to the other end of the 6.2744.040
coupling.

Use a 6.2744.090 long pressure screw to connect the capillary of
the MSM-HC labeled with Detector to the input of the MCS (see
Chapter 3.20.2, page 63)
.

Connect the detector inlet capillary with a 6.2744.090 long pres-
sure screw to the output of the MCS (see Chapter 3.20.2, page
63)
.

4 Installing the sample path

NOTE

The sample degasser does not have to be connected. We recom-
mend the usage of the sample degasser only if the sample matrix
requires it.

Connect the 6.1803.040 sample aspirating capillary (3-8) connec-
ted at the sample input of the injection valve to the output of the
sample degasser with the aid of a 6.2744.090 long pressure
screw (see Chapter 3.15, page 48).

Connect a part of the 6.1803.040 sample aspirating capillary with
a 6.2744.090 long pressure screw to the input of the sample
degasser. Guide the other end out of the instrument through a
capillary feed-through.