Figure 50, Msm-hc – components – Metrohm 850 Professional IC Anion MSM-HC MCS LP-Gradient User Manual
Page 108

5.11 High Capacity Metrohm Suppressor Module (MSM-HC)
■■■■■■■■■■■■■■■■■■■■■■
98
■■■■■■■■
850 Professional IC–Anion – MSM-HC – MCS – LP Gradient
1
2
3
4
5
Figure 50
MSM-HC – Components
1
Union nut
2
MSM-HC connecting piece 6.2835.010
3
MSM-HC rotor A 6.2842.000
4
MSM-HC housing
5
Slot in the MSM-HC housing
Cleaning the MSM-HC
Clean the MSM-HC as follows:
1 Disconnecting the MSM-HC from the IC system
■
Switch off the instrument.
■
Disconnect the MSM-HC from the separation column, peristaltic
pump and detector.
2 Dismantling the MSM-HC
■
Unscrew union nut (50-1) from the MSM-HC housing (50-4).
■
Pull the MSM-HC connecting piece (50-2) and the MSM-HC rotor
A (50-3) out of the MSM-HC housing (50-4). The connecting
piece and the MSM-HC rotor A normally stick to one another – if
this is not the case: Take a sharp object, insert into the slot (50-5)
in the MSM-HC housing, and pull out the MSM-HC rotor A (50-3)
in this way.
■
Detach the MSM-HC connecting piece (50-2) from MSM-HC rotor
A (50-3).
3 Cleaning the supply and discharge lines
■
Connect in turn each of the 6 capillary tubings fastened on the
MSM-HC connecting piece (50-2) on the high pressure pump and
pump through ultra pure water.