Accuracy, Bayer accuracy evaluation, High control solution – Bayer HealthCare Ascensia Contour Blood Glucose Monitoring System User Manual
Page 28: Normal control solution

24
Accuracy
1. Bayer Accuracy Evaluation:
To assess the accuracy of the Ascensia C
ONTOUR
System, 60
fresh fingerstick specimens from people with diabetes were
tested with the Ascensia C
ONTOUR
System (12 – 14 readings
per fingerstick) and the YSI 2300 STAT Plus Glucose Analyzer.
The glucose concentration of the samples ranged from 45 to
454 mg/dL with an average of 150 mg/dL. The range of
hematocrits was 32% to 49% with an average of 40%.
High Control Solution
Duplicate
Overall Overall
Reagent
Assays
Mean
SD
CV
Lot
Operator
(n = )
(mg/dL)
(mg/dL)
(%)
22B
HCP
(n = 1)
35
337.2
13.28
3.9
Lay User
(n = 51)
76
341.5
14.94
4.4
22C
HCP
(n = 4)
39
351.9
8.98
2.6
Lay User
(n = 53)
78
351.2
11.16
3.2
23B
HCP
(n = 5)
74
345.6
10.41
3.0
Lay User
(n = 104)
154
324.8
9.44
2.8
Normal Control Solution
Duplicate
Overall Overall
Reagent
Assays
Mean
SD
CV
Lot
Operator
(n = )
(mg/dL)
(mg/dL)
(%)
22B
HCP
(n = 1)
35
119.6
4.58
3.8
Lay User
(n = 51)
76
120.6
3.90
3.2
22C
HCP
(n = 4)
39
122.0
3.23
2.6
Lay User
(n = 53)
78
122.3
4.88
4.0
23B
HCP
(n = 5)
74
120.9
3.04
2.5
Lay User
(n = 104)
154
120.1
3.67
3.1