Bird Technologies 5010 Series-Manual User Manual
Page 24

12
Uncertainty
b
± 8% of full-scale peak
envelope power (95%
c.l.)
Match Measurement
Match Range
Return Loss
Rho (
ρ
)
VSWR
0 to 20 dB
0.1 to 1
1.22 to 99.99
Uncertainty
Twice the Avg Power
Uncertainty
(Calculated from
forward and reflected
uncertainty)
Settling Time, Max
2.5 seconds
Impedance, Nominal
50 ohms
Insertion Loss, Max
0.05 dB up to 1 GHz
Input VSWR, Max.
1.05:1 up to 1 GHz
Directivity, Typicala
30 dB
RF Connectors
QC Type (N(F) normally
supplied)
Power Supply
From host instrument
via cable
Mechanical Shock and Vibration
In accordance with MIL-
PRF-28800F Class 3
CE
CE compliant. Refer to
DOC for specific
standards.
Recommended Calibration
Interval
1 year