Installing thermoswitch – Bird Technologies 8865SC13-Manual User Manual
Page 16
Bird 8860 Series Semiconductor Termaline Coaxial Load Resistor
6
Installing
Thermoswitch
Bird 8860 series loads can be equipped with an optional interlock
thermoswitch, P/N 8890-008. It is normally closed, opening at 236 °C
(457 °F), with a rating of 10A @ 120Vac and 5A @ 230Vac.
To install or replace the thermoswitch:
1. Remove the vent plug and install the shipping plug.
2. Stand the unit on its front, supporting it so that the connector is
not damaged. In this position there is no danger of the coolant
pouring out through the socket plug hole.
3. Remove the socket plug just above the connector assembly, using
a
9
⁄
16
" hex wrench.
4. Replace the plug with the thermoswitch. Sparingly apply pipe
sealing compound to the external threads, only, of the
thermoswitch. Do not contaminate the coolant with pipe sealant.
5. Check for coolant leaks upon completion.
WARNING
The vent plug must be used at all times when the unit is
operating or cooling. Failure to do so could result in an
explosion or severe burns.