39

Operation Method

Set the Operation Stop Signal Input Mode

1

Select the item in function menu
• Check that the power in turned on.

FUNCTION

TROUBLE

ENTER

CANCEL

Stop signal

OFF



Press the FUNCTION key. The function menu

starts and the Sub Display displays the items.
Select the item using the FUNCTION key.



Display the "Stop signal" and press the ENTER

key.

™

Press the CANCEL key to cancel the function menu.

2

Select mode

Stop signal

OFF

Stop signal

ON


The "" (cursor) goes out and the mode currently

selected blinks instead. Display the "ON" or
"OFF" using the "▼▲".

3

Set operation stop signal input mode

TROUBLE

ENTER

CANCEL

Stop signal

ON


Press the ENTER key. The operation stop

signal input mode is decided and the function
menu selection screen is displayed.

• The unit monitors the input of operation stop signal

during operation in each mode and automatically
stops operation if it is input when the operation
stop signal input mode is set to "ON". The unit
enters to the standby state.