Operation method, Auto stop operation, Set temperature and operation period/stop time – Yamato Scientific IN603W Low Temperature Incubators User Manual
Page 22
![background image](/manuals/576638/22/background.png)
18
Operation Method
Auto Stop Operation
3
Set temperature and operation period/stop
time.
AUTO START
FIXED TEMP
AUTO STOP
PROGRAM
STANDBY
END
HEATER
FUNCTION
PROGRAM
MENU
ENTER
CANCEL
POWER
TROUBLE
℃
Set TEMP
0.0℃
Auto-stop OPR
REFRIGE
DEFROST
AUTO START
FIXED TEMP
AUTO STOP
PROGRAM
STANDBY
END
HEATER
FUNCTION
PROGRAM
MENU
ENTER
CANCEL
POWER
TROUBLE
℃
Set TEMP
37.0℃
Auto-stop OPR
REFRIGE
DEFROST
①
The setting temperature input screen is
displayed. The Sub Display indicates "Set
TEMP" and the numeric character that indicates
temperature blinks.
②
Set the temperature using the "▲▼".
③
Press the ENTER key to decide the temperature.
④
The operation period/stop time input screen is
displayed after the setting temperature is
decided.
Display the period/time using the "▲▼".
Input the operation period when the setting of
timer mode shows "Time". Input the operation
stop time when it shows "Clock".
OPR time
30min
Auto-stop OPR
Stop time
13:00
Auto-stop OPR
(Operation period
edition screen)
(Operation stop time
edition screen)
• The display style of operation period varies
depending on the range of time to be displayed.
Time Range
Indication
0minute to 59minutes
0min to 59min
1hour to
99hours59minutes
1h00m to 99h59m
• The input range of operation stop time is always
from 0:00 to 23:59.
Pressing the CANCEL key quit the setting of auto
stop operation.