26

Operation Method

Input Program

2


• The temperature is kept constant till the end of setting time when the "START TEMP" and the "Set

Temp" are the same. The unit enters into "Wait" and measurement of residual time is suspended
when the temperature in furnace is within ±1℃ to the setting temperature.

Set Temp.

TEMP.

TIME

Set Time

TEMP

.

START

.


• The unit operates with full-power from the "START TEMP" to the "Set Temp" when the period is set

to 0 minute. It keeps the "Wait" state until the temperature in furnace is within ±1℃ to the
setting temperature.

Set Temp.

TEMP.

TIME

Reached time to the set temp.

TEMP

.

START

.


• The conception of repeat is shown in the figure below. The first operation of repeat interval is not

counted as a repeat count.

Step No. S04

Start step: S02
Repeat times: 5

Step No. S01

No repeat

instruction

Step No. S02

No repeat

instruction

Step No. S03

No repeat

instruction

Step No. S05

No repeat

instruction

5 repeating

of interval

Waiting