Operation Method

Auto Stop Operation

Auto stop operation
procedure

This operation is used to specify the automatic stop time in the
fixed temperature operation.

1. Set stop time

① Press the TIMER key on the initial screen.

Press the TIMER key again. The setting temperature display
screen displays the character "AstP", which indicates the auto
stop operation, with blinking.

② Press the ENTER key.

The measurement temperature screen displays the character
"SV", which indicates the temperature setting. The setting
temperature screen displays the current setting temperature with
blinking. The AUTO STOP lamp blinks, too.

③ Set the temperature using the "▼▲".
④ Press the ENTER key again.

The measurement temperature display screen displays the
character "tim", which indicates the timer setting. The setting
temperature display screen displays the current setting time with
blinking.

⑤ Set the time using the "▼▲".

Timer function:

• The maximum setting time is "999 hours and 50 minutes".
• The time can be set in increments of a minute under 99 hours

and 59 minutes.

• It can be set in increment of ten minutes over 100 hours.
• The

"▼▲"can change the setting time quickly when it is pressed

continuously. Press them discontinuously when fine adjustment
is needed.

2. Start timer operation

• Press the START/STOP key for one second after deciding the

time.

• Timer operation starts with the AUTO STOP lamp lighting on.
• The timer is activated at the point when the temperature in

furnace (measurement temperature) reaches to the setting
temperature.