beautypg.com

Specification – Yamato Scientific IJ100 Low Temperature Incubator User Manual

Page 23

background image

21

Specification

Model IJ100

Operating temperature range

0 to 60

℃ (room temp: 20℃, no load)*

Temperature adjustment accuracy

±0.3℃ (the center of inner bath, set temp.:37℃)

Temperature distribution accuracy

±1.0℃ (the center of inner bath, set temp.:37℃)

Temperature rising time

Approx. 40min. (20 to 60°C)

Performance

Temperature falling time

Approx. 100min. (20 to 0°C)

Heating device

Mica heater 120W

Cooling device

Two peltier cells, forced radiating method:
It works continuous at the setting temperature below 35

℃.

(It is turned off at the setting temperature 35

℃ and over.)

C

onfiguration

Internal fan

Axial fan (DC)

Controller Hitec

IV

FR

type1

Temperature control system

PID control of heater output by microcomputer

Temperature setting system

Digital Setting by

▼▲ (up/down) Keys

Operational mode

Fixed temperature operation

Sensor Thermistor

C

o

ntrol Section

Additional function

Temperature preset function (1 for desired temperature is able to
set/load)

Self-diagnostic function

Sensor error, Heater error, Triac error, Main relay error, Auto
overheating prevention

Sa

fe

ty

D

e

vi

ce

Preservation device

Earth leakage breaker, Independent overheating prevention
device

External dimensions

350(W)

×396(D)×535(H) mm

Internal dimensions

250(W)

×250(D)×250(H) mm

(Air jacket installed: 200

Ч200Ч200 mm)

Internal capacity

15.6L

Door

Single door with magnet packing

(the direction of opening/closing is changeable)

Weight Approx.

20kg

S

tandard

Power supply

100V AC 50/60Hz 3.5A

Accessories

Shelf

× 2 (Load: 5kg for each), Plate for drainage, Instruction

manual

Optional accessories

Extra shelf (with metal fittings), Air jacket made of aluminum (with
shelf)

The values shown on “Performance” are in power supply 100V AC.
Usable ambient temperature for this unit is 5 to 30℃.