2. Before operating the unit

Precautions when installing the unit

10. Always operate the unit with the vent holes open.

Do not cover the vent holes on the top panel of the unit.
Adjust the open amount according to the water content of a specific sample.

11. Be sure to connect the power plug to the dedicated power distribution panel or

a wall outlet.

Use a power distribution panel or a wall outlet that meets the electrical capacity of the unit.

Electrical
capacity:

DX302 AC100V 9.5A
DX402 AC100V 14A
DX602 AC100V 14A

When the unit will not start even when you turn the earth leakage breaker to “ON”, check for
low main voltage or if the unit is connected to the same power supply line as other devices
and connect it to another line if necessary.

Avoid connecting too many devices using a branching outlet or extending a wire with a cord reel
or temperature controlling function may degrade due to voltage drop.

Do not connect the unit to any parts or lines other than a correct power supply line such as a
gas pipe, a water pipe or a telephone line.
Otherwise, an accident or a malfunction may result.