beautypg.com

Operating procedures – Yamato Scientific DX602 Constant Temperature Drying Oven User Manual

Page 15

background image

13

4. Operating procedures

List of operation modes and functions


Functions of the unit are as shown below:

Name

Description

Page

1

Overheat prevention
function

Automatic overheat prevention function:

This function is linked to the unit set temperature and has
been set to so that it is automatically activated (returned
automatically) at a temperature 12℃ higher than the set
temperature in the bath.

Standalone overheat protection device (Thermostat):
When the temperature in the bath reaches the set

temperature of the standalone overheat protection device
(Thermostat), its heater circuit trips to shut off controller
operation.
The temperature can be set with the manual dial on the
hydraulic overheat prevention device installed at the right
side of the unit.

P.15

2

Calibration

Offset function

Calibration offset function compensates any differences
between the target temperature in the bath and the control
temperature of the controller (sensor temperature.)
The function can compensate to either plus or minus side
for the whole temperature band of the unit.
This compensation can be set with the SUB MENU keys.

P.23

3 Setting

lock

function

This function locks the set operation status.
The lock can be set or released with the SUB MENU key.

P.24

4

Power outage
compensation function

This function returns the main unit operation to the resume
status after recovery from power outage, or keeps the
current stop status.
This compensation can be set with the SUB MENU keys.

P.25

This manual is related to the following products: