4. Operating procedures
Operating procedures (fixed temperature operation)
How to start fixed
temperature operation
MEASURED TEMP.
℃
HEATER
RUN
設定温度
1.Turn the ELB ON. (Turn the ELB to “ON.”)
When the ELB is turned ON, the intial values will be
displayed for about four seconds, then the initial screen
will appear and the current bath temperature and the
previous set temperature are displayed on each of the
indicators.
Measured temperature screen: Displays the current bath
temperature
Set temperature screen: Displays the previous set
temperature
MEASURED TEMP.
℃
HEATER
RUN
SET TEMP.
RUN
STOP
TIMER
SUB MENU
2. Setting the temperature
Set a temperature using the ▼▲ keys.
MEASURED TEMP.
℃
HEATER
RUN
SET TEMP.
RUN
STOP
TIMER
SUB MENU
3. Starting operation
Press the RUN/STOP key longer.
Fixed value operation will start and the RUN lamp and the
HEATER lamp come on.
4. Stopping operation
Press the RUN/STOP key longer.
Operation stops, the RUN lamp goes off and the screen
switches to the initial setting screen.
When you want to correct
setting errors or change
settings
When you want to change settings, press the ▼▲ keys on
the current screen to enter the setting mode where you can
change settings. Blink stops three seconds after three
seconds after change and setting is completed.
Caution
①
When you want to lower the set temperature during fixed temperature operation,
note that it takes some time to reach the reset temperature since the unit has no
cooling capacity.
②
Immediately after operation has been stopped, the temperature in the bath is
around the set temperature. Operation stop refers only to machine stop and time
needed for decreasing the emperature in the bath is not considered.