4. Operating procedures

Operating procedures (auto stop operation)


This mode automatically stops fixed temperature operation after a certain time from its start set with
the timer.

Procedures for auto stop
operation

MEASURED TEMP.

HEATER

RUN

SET TEMP.

RUN
STOP

TIMER

SUB MENU

1. Setting a stop time

After confirming the temperature you want is set,

Press the TIMER key to display characters
AStP

on the measured temperature screen that

indicate auto stop operation.
The set time is displayed on the set temperature
screen.

Set a time you want using the ▼▲ keys.

Pressing the▼▲ keys makes the set time blink. The

time is determined when blinking stops.

About the timer function

The maximum time that can be set for the timer is 999 hours
50 minutes.
Up to 99 hours 59 minutes, time can be set in minutes.
One hundred hours and over are set only in 10 minutes.
Keep the ▼▲ keys pressed to continuously change set time
and you can quickly reach the time you want. Press the ▼

keys once at a time for fine adjustment.

MEASURED TEMP.

HEATER

RUN

SET TEMP.

RUN
STOP

TIMER

SUB MENU

2. Starting timer operation
When the time you want is set, press the RUN/STOP key

for about one second while characters AStP

that

indicate auto stop operation are displayed on the
measured temperature screen and the set time on the set
temperature screen.

The RUN lamp blinks and timer operation is started.
Timer starts counting when the temperature in the bath

reaches the set temperature.

Once timer counting is started, the set temperature screen

changes to the remaining time display.

MEASURED TEMP.

HEATER

RUN

SET TEMP.

RUN
STOP

TIMER

SUB MENU

3. Stopping and ending timer operation
Operation stops automatically when the set temperature

has elapsed.

Characters End

blink on the set temperature

screen to indicate operation has ended.
Press the RUN/STOP key for approx. one second to end
the timer operation mode. The screen switches to the initial
setting screen.