7

Installation

Do not use the unit in an area where there is flammable or explosive gas.

Never use the unit in an area where there is flammable or explosive gas. The unit is
not explosion-proof. An arc may be generated when the power switch is turned ON
or OFF, and fire/explosion may result.



Explosive gas

Flammable gas


Always ground the unit.

z Connect the grounding cable to your ground conductor or ground terminal.

z Do not forget to ground the Oven, to protect you and the unit from electrical shock

in case of power surge.

z Do not connect the grounding wire to a gas pipe, or by means of a lightning rod or

telephone line. A fire or electrical shock will occur.

NOTE: The oven has not a plug connected because of the 200V specification.

Choose a correct power distribution board.

z Choose a correct power distribution switchboard or receptacle that meets the

oven’s rated electric capacity.

Electric capacity

AC 200 V, single phase 16A

z Do not connect the oven to an outlet that differs from the above specifications

because a fire or electrical shock will occur.

Supply connections for the oven

z Request the supply connection for the 200V specifications from a licensed

electrician.

z Failure to have this operation complete by certified personnel will cause a fire or

electrical shock during Oven operation.


Install the Oven on a level area.

z Do not installation the oven on a non level surface. This will cause hazards to the

operator and create problems during actual operation.