Description/specifications, Operation, Maintenance, repair and technical assistance – UVP High Performance Ultraviolet Transilluminators User Manual
Page 2: Changing the replacement tubes, Cleaning the transilluminator
Description/Specifications
The dimensions of the High Performance UV Transilluminators are:
Width:
19.13” (486mm)
Depth:13.25” (336mm)
Height: 5.63” (143mm) Height in-
cludes cover
Operation
1. Place the transilluminator on a level work surface allowing air space around the bottom of the unit
and the work surface. This space allows for the proper air circulation through the unit.
2. Plug the female end of the power cord into the transilluminator. For 230V models, or those requir-
ing special power cord connectors, ensure that the proper configuration of male connector or plug
has been properly connected to the power cord.
3. Plug the male end of the power cord into a properly grounded electrical outlet. The proper voltage
of the transilluminator is found on the product information label.
4. The transilluminator comes equipped with a UV Blocking Cover. If this cover is not in place or has
been removed, do not operate the unit without securing the cover. If the cover is missing, a UV
Blocking Faceshield must be worn to avoid UV exposure to the skin. UV Blocking Eyewear should
be worn even with the cover in place to avoid accidental UV exposure. The cover is reversible and
can open away from the researcher by removing the four screws (two on both ends of the transil-
luminator) and rotating the transilluminator top 180
o
.
5. Place gel/sample on the filter area. Gels may be placed on a UVP Gel-Tray (see Section 8 for part
numbers) which protects the filter surface from cuts and scratches. It is recommended that gloves
be worn to avoid contact with gel and staining agents.
6. Position the Dual Intensity Switch to the High operation position.
7. Press the ON/OFF switch to ON. The UV tubes within the unit should be glowing beneath the filter
after a momentary flickering during the start-up period.
Part
Wave-
# of
Intensity
Model
Number
length
Volt/Hz/Amp
Filter Size
Tubes
Style
TFM-20
95-0286-01
302nm
115V/60Hz/2.0
20 x 20 cm
4 x 25W
High/Low
TFM-20
95-0286-02
302nm
230V/50Hz/2.0
20 x 20 cm
4 x 25W
High/Low
TFM-20
95-0286-03
302nm
100V/50-60Hz/2.0
20 x 20 cm
4 x 25W
High/Low
TFM-26
95-0285-01
302nm
115V/60Hz/2.0
21 x 26 cm
4 x 25W
High/Low
TFM-26
95-0285-02
302nm
230V/50Hz/2.0
21 x 26 cm
4 x 25W
High/Low
TFM-26
95-0285-03
302nm
100V/50-60Hz/2.0
21 x 26 cm
4 x 25W
High/Low
TFM-30
95-0302-01
302nm
115V/60Hz/2.0
25 x 30 cm
4 x 25W
High/Low
TFM-30
95-0302-02
302nm
230V/50Hz/2.0
25 x 30 cm
4 x 25W
High/Low
TFM-30
95-0302-03
302nm
100V/50-60Hz/2.0
25 x 30 cm
4 x 25W
High/Low
TFM-40
95-0283-04
302nm
115V/60Hz/2.0
20 x 40 cm
4 x 25W
High/Low
TFM-40
95-0283-05
302nm
230V/50Hz/2.0
20 x 40 cm
4 x 25W
High/Low
TFM-40
95-0283-06
302nm
100V/50-60Hz/2.0
20 x 40 cm
4 x 25W
High/Low
TFL-40
95-0283-01
365nm
115V/60Hz/2.0
20 x 40 cm
4 x 25W
High/Low
TFL-40
95-0283-02
365nm
230V/50Hz/2.0
20 x 40 cm
4 x 25W
High/Low
TFL-40
95-0283-03
365nm
1005V/50-60Hz/2.0
20 x 40 cm
4 x 25W
High/Low
2
8. To lower the UV intensity for extended viewing or sample work, simply press the Dual Intensity
Switch to low. A drop in the intensity will now occur. The fluorescence of the stained sample at this
time should diminish. As UV radiation is not visible to the eye, the UV transmitting filter does not
allow the passage of light up to 400nm (blue). A slightly dimming of this blue color is visible at this
time. A UV measuring device with the proper UV wavelength sensor can be used to provide actual
intensity reading (see Section 8 for Radiometer information).
9. After viewing/photographing the sample, turn the unit off.
10. Clean the surface with a damp soft cloth or sponge. Never use abrasive cleaners (can damage the
UV filter surface).
Maintenance, Repair and Technical Assistance
UVP offers technical support for all of its products. If you have any questions about the product’s use,
operation or repair, please call or fax UVP Customer Service in the US at (800) 452-6788 or (909) 946-
3197, Fax: (909) 946-3597; In Europe/UK: +44(0)1223-420022; Fax: +44(0)1223-420561.
Note: A Returned Goods Authorization (RGA) number must be obtained from UVP Customer
Service before returning any product.
Changing the Replacement Tubes
3
1. Disconnect the transilluminator from the electrical supply.
2. A Phillips head screwdriver is required to remove the filter cover.
3. Carefully twist the UV tubes from their sockets.
4. Fit with the proper replacement tubes (refer to Section 9 for ordering information).
Cleaning the Transilluminator
The painted surfaces and filter areas of the transilluminator should be cleaned with water, soap, and a
sponge or cloth towel. Never use abrasive cleaners, solvent-based cleaners or scouring pads. Always
disconnect the transilluminator from the electrical power prior to cleaning.