Hybridization Ovens
Page 12
call UVP.
11. This calibration process should be repeated following any unsuccessful recalibration. If calibration is
unsuccessful, replace the new UV wavelength tubes with the previously-removed old wavelength tubes.
12. Successful recalibration to the NEW UV wavelength changes the values in the microprocessor and the
numeric setting value is displayed on the LCD. To return to original or another wavelength, the proper
UV calibration sensor is needed.