Magnetic wand, Gas detector – RKI Instruments M2A User Manual

Page 10

background image

5

M2A Transmitter Operator’s Manual

Use the junction box’s two mounting holes to mount the M2A to a vertical surface at the
monitoring site. The window in the cover on the front of the junction box allows you to view the
OLED display and use the magnetic wand to actuate the magnetic control switches so you can
perform non-intrusive calibration. Removing the cover allows you to access the interior of the
junction box.

Magnetic Wand

The magnetic wand is a short plastic rod with a magnet in one end. It is used to actuate the
magnetic control switches on the control PCB while the junction box cover is still installed so that
non-intrusive calibration can be performed.

Gas Detector

The gas detector senses the target gas and is mounted in a 3/4” conduit hub on the right bottom side
of the M2A. A variety of detectors may be used with the M2A. Table 2 below lists the detectors that
can be used with the M2A.

Table 2: M2A Detectors

Target Gas, Range

Detector(s) Used

Combustible Gas, 0 - 100% LEL

61-0140RK (Catalytic UL version)

61-0140RK-05 (Catalytic CSA version)

Hydrogen, 0 - 100% LEL

NC-6205-01 (Catalytic UL version, hydrogen specific)

NC-6205-05 (Catalytic CSA version, hydrogen specific)

CH

4

, 0 - 100% LEL

61-0190RK-CH4 (IR UL)

HC, 0 - 100% LEL (cal. to propane)

61-0190RK-HC (IR UL)

CH

4

, 0 - 100% volume

61-0192RK-CH4 (IR UL)

CO

2

, 0 - 5,000 ppm

61-0191RK-02 (IR UL)

61-0198RK-02 (IR)

CO

2

, 0 - 5%

61-0191RK-03 (IR UL)

61-0198RK-03 (IR)

CO

2

, 0 - 50%

61-0191RK-05 (IR UL)

61-0198RK-05 (IR)

CO

2

, 0 - 100%

61-0191RK-10 (IR UL)

61-0198RK-10 (IR)

Oxygen, 0 - 25%

65-2494RK (capillary)

65-2510RK (partial pressure)

65-2511RK (capillary)

65-2514RK (capillary, CSA)

CO, 0 - 300 ppm

65-2433RK-05 (CSA)

65-2496RK

H

2

S, 0 - 100 ppm

65-2423RK-05 (CSA)

65-2495RK

NH

3

, 0 - 75 ppm

65-2300RK-NH3

AsH

3

, 0 - 1.5 ppm

65-2300RK-ASH3