RKI Instruments SDM-2009 User Manual
Page 37
Calibrating a GX-2009 • 33
hold the EDIT ENTER and POWER buttons simultaneously for about one
second and then release them. To avoid accidentally entering Edit Mode,
press and hold the POWER button first, then press and hold the EDIT
ENTER button.
CAUTION:
When using the GX-2009 with the SDM-2009, do not turn off the
instrument using the instrument power button. Use the EDIT ENTER
and POWER buttons on the SDM-2009 to turn off the instrument.
The CAL
S
LED will remain on indicating the test results. If the same
GX-2009 is turned on again, the test results will still be indicated by the
CAL
S
LED and on the GX-2009 screen. To clear the CAL
S
LED, with the
GX-2009 on, press and hold the EDIT ENTER button for about 3 seconds
until the display returns to the home screen. If a new GX-2009 is turned on
and connected, the results displayed by the CAL
S
LED will automatically
be cleared.
17. The results of the calibration will be stored in the SDM-2009’s memory and will
be available to copy to a USB flash drive. See “Copying Calibration and Bump
Test Records to a USB Flash Drive” on page 38 for instructions to copy the
saved bump test and calibration records to a USB flash drive.
18. Remove the GX-2009 from the SDM-2009.
19. If you wish to calibrate additional instruments, repeat step 5 - step 18 above
for each additional instrument.