GX-2003 Operator’s Manual

Parts List • 86

6. If necessary, use the ▲AIR and (SHIFT)▼ buttons to adjust the

reading to match the concentration listed on the calibration
cylinder.

7. Press and release the POWER ENTER button to save the span

value. The LCD will indicate that the calibration has ended, then
the single calibration menu displays.

8. Disconnect the tubing from the GX-2003’s probe.

9. Repeat steps 3 through 8 for any other channels you want to

calibrate. Make sure you use an appropriate calibration cylinder
for each channel.

CAUTION:

When calibrating the oxygen channel, verify the concentration
of oxygen listed on the cylinder’s label. For oxygen-free samples
(100% nitrogen for example), set the oxygen zero setting to
0.0%.

10. After the last channel is calibrated, disconnect the calibration

tubing from the probe.

11. With the Single Calibration menu displayed, press the (SHIFT)▼

button until the cursor is next to the

ESCAPE

menu item.

12. Press and release the POWER ENTER button to return to the

calibration menu.

13. Use the (SHIFT)▼ button to place the cursor next to the NORMAL

OPERATION

menu item, then press and release the POWER

ENTER button to return to Measuring Mode.

Parts List

Table 8: Sample Bag Calibration Kit Spare Parts

Part Number

Description

81-0016RK-05

Calibration cylinder, 100% volume methane in nitrogen, 58 liter

81-0018RK-01

Calibration cylinder, 50% isobutane in air, 34 liter

81-0090RK-01

Calibration cylinder, 3-gas mix, LEL/Oxygen/CO, 34 liter

81-0154RK-04

Calibraition cylinder, 4-gas mix, LEL/Oxygen/CO/H2S, 34 liter

81-1001RK

Dispensing valve, for 17/34 liter cylinders

81-1051RK

Regulator with gauge and knob, 34AL/58 liter/103 liter cylinders,0.5 LPM

81-1126RK

Gas bag with clamp and hose barb, 9” x 9”, 2 liter