beautypg.com

Preparing for calibration – RKI Instruments GX-2003 Manual User Manual

Page 88

background image

81 • Preparing for Calibration

GX-2003 Operator’s Manual

To calibrate the GX-2003, you will need:

• Known calibrating samples of the gases being detected. The

combustible and toxic gas samples should have concentrations in
approximately the middle of the range of detection. An oxygen-
free source, such as 100% nitrogen is recommended for setting the
oxygen zero when using single calibration.

WARNING: When using auto calibration, the multi-gas cylinder

must have an oxygen concentration in the range of 10% -
18% oxygen.

• A gas collection bag with hose clamp

• A 0.5 LPM fixed-flow regulator or dispensing valve

• Calibration tubing

To calibrate the %LEL, Oxygen, CO, and H

2

S sensors at the same time,

automatically, with no need for a zero-oxygen source, you can use the
auto calibration feature with a 4-gas cylinder. If the H

2

S channel is not

active, then a 3-gas cylinder may be used for auto-calibration. This
document includes instructions for auto calibration with a fixed flow
regulator or dispensing valve and a 4-gas cylinder. This document
also includes instructions for calibrating one channel at a time using
single calibration.

Preparing for Calibration

Whether you will calibrate the Model GX-2003 using auto calibration
or single calibration, you will need to set the fresh air readings first
before setting the span readings for each gas.

1. Find a fresh-air environment. This is an environment free of toxic

or combustible gases and of normal oxygen content (20.9%).

2. Turn on the unit as described in “Turning on the GX-2003” in the

GX-2003 Operator’s Manual using the sample hose and probe.

3. With the unit on, perform a demand zero by pressing and holding

the ▲AIR button. The display prompts you to hold the ▲AIR
button.

4. Continue to hold the ▲AIR button until the display prompts you

to release the ▲AIR button. The GX-2003 will set the fresh air
reading for all channels.