61 • Using Setup Mode

GX-2003 Operator’s Manual

2. Use the ▲AIR or (SHIFT) ▼ button to display the desired

character, then press and release the POWER/ENTER button to
enter the character. The next character flashes.

3. Repeat step 2 to enter the remaining characters. The serial number

is saved and the main menu displays after you enter the last
character.

Turning the User/Station ID Function On or Off

With USER/STATION ID set to ON, the ID Screen displays during
start up and in Display Mode. The ID’s can be updated in Display
Mode.

With USER/STATION ID set to OFF (factory setting), the ID Screen
does not display during start up or in Display Mode.

1. From the main menu, place the cursor next to the USER/

STATION ID

menu item. Press and release the POWER ENTER

button.

2. Use the ▲AIR or (SHIFT) ▼ button to display the desired setting.

3. Press and release the POWER ENTER button to save the setting

and return to the main menu.

Calibrating Using Auto Calibration

If the CAL. EXPIRED ACTION is set to CANNOT USE and
calibration becomes due, then the unit must be calibrated using this
menu item and/or the next menu item, SINGLE CALIBRATION
because it will not be possible to enter normal operation. See
“Calibration Supplies and Equipment” on page 32 for the items you
will need to perform a calibration.

1. Install the demand flow regulator onto the calibration cylinder.

2. Connect the sample tubing to the demand flow regulator.