Overview • 14

Chapter 3: Using the GX-2003 in Normal Mode

Overview

This chapter explains how to operate the GX-2003 in Normal Mode.
Normal Mode is used to perform confined space entry monitoring or
general area monitoring. While in Normal Mode, Display Mode and
Calibration Mode are accessible.

If a standard version of the GX-2003 is ordered, it is shipped with Bar
Hole Mode and Leak Check Mode disabled so that the instrument
only runs in Normal Mode when turned on.

Start Up, Normal Mode

This section explains how to start up the GX-2003 in Normal Mode,
get it ready for operation, and turn it off

NOTE:

The screens illustrated in this section are for a 4-gas unit that
has the % LEL range for the CH

4

channel. The screens

displayed by your GX-2003 may be slightly different.

Turning On the GX-2003, Normal Mode Only

The following description of the GX-2003 start up sequence assumes
that the following menu items in Setup Mode are turned on: LUNCH
BREAK, USER/STATION ID, and CAL. TIME REMAINING. If any of
these items is turned off, then the corresponding screen will not
appear. This description also assumes that the LB Mode item in Setup
Mode is turned off so that Bar Hole Mode and Leak Check Mode are
not active.

1. Connect the tapered rubber nozzle or the sample hose to the GX-

2003’s quick connect inlet fitting.

2. If a sample hose is used, screw the probe onto the sample hose’s

threaded fitting.

3. Press and briefly hold down the POWER/ENTER button. Release

the button when you hear a beep.