44 • Calibration

Eagle Instruction Manual

2. Press the POWER/ENTER button to display the Calibration

Values screen.

The gas concentrations displayed in the Calibration Values screen
must match the gas concentrations listed on the Four-Gas Calibration
Cylinder. If all concentrations match, go to step 7. If one or more
concentrations do not match, continue with step 3.

3. To adjust the values on the screen, hold down the SHIFT/▼

button, and press the DISP/ADJ button. The Auto Calibration
screen for the combustible gas channel displays.

4. Use the AIR/▲ (increase) and SHIFT/▼ (decrease) buttons to set

the correct combustible gas value.

5. Press the POWER/ENTER button to enter the new setting. The

Auto Calibration screen for the next channel displays.

6. Repeat steps 4 and 5 to set the correct values for the remaining

channels and return to the Calibration Values screen.

NOTE:

The RKI Four Gas Cylinder contains approximately 12% O

2

by volume. Be sure to set the “OXY” reading to agree with the
concentration listed on the cylinder’s label, not zero.

7. With the Calibration Values screen displayed, press the

POWER/ENTER button. The gas readings flash.

8. Connect the tubing from the regulator to the Eagle’s probe.Wait

approximately 1 minute or until the readings stabilize.

9. Press the POWER/ENTER button to set the calibration to the

programmed values.

If a sensor(s) cannot calibrate to the proper value,

FAIL PUSH AIR

KEY

displays and the Eagle lists the sensor(s) that failed to

calibrate. (The other sensors calibrate normally.) The buzzer and
alarm lights activate. Press the AIR/▼ button to reset the alarm