beautypg.com

Hose and probe, Sensors – RKI Instruments Inert Gas Testing Eagle with Two Pumps User Manual

Page 13

background image

Eagle Instruction Manual

Description • 7

Hose and Probe

A 5 foot polyurethane sample hose and a 10 inch hydrophobic probe
are included as standard. The hose has a male quick connect fitting on
one end and a female quick connect fitting on the other end. The
probe has a male quick connect fitting. Normally, the male end of
sample hose is installed in the Eagle inlet fitting and the probe is
installed in the female end of the hose. However, if the sample hose is
not needed for monitoring a particular area, the probe may be
installed directly to the inlet fitting. Sample hose lengths are available
from 5 feet (standard length) to 125 feet (see “Appendix A: Parts List”
on page 55).

CAUTION:

Sample hose lengths of more than 125 feet are not recommended
for the Eagle because of flow rate reduction and increased
response time. Consult RKI Instruments, Inc. for hose lengths
longer than 125 feet.

The standard probe includes a replaceable particle filter and
hydrophobic filter disk that prevent particulates and water from
entering the Eagle’s flow system. See “Replacing the Probe’s Particle
Filter and Hydrophobic Filter Disk” on page 49 for instructions to
replace the particle filter and hydrophobic filter disk.

Sensors

This section describes the Eagle’s sensors. Your specific Eagle model
may not include all of the sensors described below. Under normal
conditions, the Eagle’s standard sensors have an operating life of

Standard

Probe

Sample Hose