beautypg.com

Alarms, Alarm indications – RKI Instruments Inert Gas Testing Eagle with Two Pumps User Manual

Page 21

background image

Eagle Instruction Manual

Alarms • 15

Alarms

Alarm Indications

This section describes the Eagle’s audible and visual alarm indications
for gas, over range, low flow, low battery, and sensor failure alarms.
This section also describes how to reset gas alarms.

The default alarm settings are listed in Table 2, “Standard Sensor
Specifications” on page 4. The alarm settings are user-adjustable as
described in “Updating the Alarm Point Settings” on page 33.

NOTE:

The screens illustrated in this section are intended as
examples only. The screens displayed by your Eagle model
may be slightly different.

First Gas Alarm

If a channel’s gas reading exceeds the first alarm setting (falls below
for the oxygen channel):

ALM1

displays in the alarm field for that channel.

• The channel’s display line flashes.

• The buzzer sounds a pulsed tone.

• The alarm lights flash.

Second Gas Alarm

If a channel’s gas reading exceeds the second alarm setting:

ALM2

displays in the alarm field for that channel.

• The channel’s display line flashes.

• The buzzer sounds a pulsed tone.

• The alarm lights flash.

C H

4

O

Y

C O

H

2

S

X

2 0 . 9

0

0

0

.

0

L

V
P
P

P
P

M
M

E

O

L
L

%
%

A L M

1

1

C H

4

O

Y

C O

H

2

S

X

2 0 . 9

0

0

0

.

0

L

V
P
P

P
P

M
M

E

O

L
L

%
%

A L M

2

5