Fault signals in the display
F02 to F13
Appearing in display
Appearing in display
F03
F04
Problem in
Open or close lid,
safety circuit of
otherwise lid lock
lid lock
defective
Service necessary!
Service necessary!
Fault appears with the lid closed if
the lid lock is defective.
Switch off machine at the
main switch and switch on
again.
If the fault reappears this is a
safety problem and servicing is
necessary.
Fault appears with the lid closed if
the lid lock is defective.
Switch off machine at the
main switch and switch
on again.
If the fault reappears this is a
safety problem and servicing is
necessary.
Appearing in display
Appearing in display
F07
F15
Motor speed control
Problem in
is defective
safety circuit of
frequency converter
Service necessary!
Appearing in display
Appearing in display
F16
F17
Motor is overheated
Motor is overheated
No START possible
Has been switched off
Please allow to cool
Please allow to cool
Continue with STOP
Appearing in display
F18
Problem in
safety circuit of
transformer
Service necessary!